Division of Biology and Medicine
BioMed Core Facilities

Zeiss Axiovert 200M Fluorescence Microscope (SFH)

Imaging of samples labeled with fluorescent indicators

Overview

BioMed Center, room BMC 007A

200MThe Zeiss Axiovert 200M Fluorescence Microscope is a motorized inverted microscope with an AxioCam MRc5 color camera and a Hamamatsu ORCA-ER monochrome camera controlled by AxioVision 4.8.2 software. This microscope is well equipped for transmitted light microscopy and fluorescence microscopy.

Operation

Objectives:

Mag

Correction

NA

Phase

DIC

WD (mm)

Coverslip #

5x

Plan-NEOFLUAR

0.15

Ph1

-

13.6

0.17mm #1 1/2

10x

Plan-NEOFLUAR

0.3

Ph1

DIC II

5.5

0.17mm #1 1/2

20x

Plan-NEOFLUAR

0.5

Ph2

DIC II

2.0

0.17mm #1 1/2

40x

Plan-NEOFLUAR

0.75

Ph2

DIC II

0.5

0.17mm #1 1/2

40x Water

C Apochromat

1.2

none

DIC III

0.28

0.17mm #1 1/2

100x Oil

Plan-NEOFLUAR

1.3

Ph3

DIC III

0.2

0.17mm #1 1/2

Theoretical XY Objective Resolution: based on: Dxy=0.61*wavelength/NA

Mag

NA

DAPI

FITC

RHO

CY5

5x

0.15

1870 nm

2175 nm

2520 nm

2930 nm

10x

0.3

935 nm

1090 nm

1260 nm

1465 nm

20x

0.5

560 nm

650 nm

755 nm

880 nm

40x

0.75

375 nm

435 nm

505 nm

585 nm

40x Water

1.2

235 nm

270 nm

315 nm

365 nm

100x Oil

1.3

215 nm

250 nm

290 nm

340 nm

Transmitted light microscopy: The Axiovert 200M has DIC and Phase Contrast on all objectives (Plan Neo 5x, 10x, 20x, 40x dry, 40x oil, and 100x oil). The high resolution color camera can be used for imaging fixed samples (e.g. histological preparations) and for time-lapse recordings of living cells. A stage heater is available for imaging living cells at a set temperature.

Fluorescence microscopy: The filter sets allow imaging of a wide variety of fluorescent indicators such as Alexa-Fluors, Calcium-Green, Cy2, Cy3, Cy5, DAPI, Fluorescein (FITC), FM1-43, Fluo-3, Fluo-4, GFP, CFP, YFP, Hoechst 33258, Rhodamine, Texas Red, TRITC, and others. The stage heater and the sensitive monochrome camera make this microscope particularly well suited for imaging dynamic patterns of fluorescence in living cells.

Filter sets on the Axiovert 200M:
1) DIC: Analyzer for DIC
2) DAPI: Chroma set 49028, Excitation 384-407 nm, Emission 435-485 nm (bandpass filter).
3) FITC: Chroma set 41001, Excitation 460-500 nm, Emission 510-560 nm (bandpass filter).
4) RHOD: Chroma set 41002c, Excitation 530-560 nm, Emission 590-650 nm (bandpass filter).
5) Cy5: Chroma set 41024, Excitation 590-650 nm, Emission 665 nm and above (long pass filter).

Also available upon request:
6) GFP: Chroma set 41017, Excitation 450-490 nm, Emission 500-550 nm (bandpass filter).
7) CFP: Chroma set 31044v2, Excitation 426-446 nm, Emission 460-500 nm (bandpass filter).
8) YFP: Chroma set 41028, Excitation 490-510 nm, Emission 520-550 nm (bandpass filter).
9) CFP/YFP FRET: Chroma set 31052, Excitation 426-446 nm, Emission 520-550 nm.

Resources

Reservations

Contact